Workflow to operate US LP Tutor study station
Before the participant arrives:
- Prepare the phantom first. The first model: PH02M01
- Turn on the Ultrasound machine and start UltraSonix software (Start Menu)
- Open Plus Server Launcher on the US machine (Start Menu, version: Plus 2.5.0.2017-05-16)
- D:/LumbarTutor/LumbarTutor/Config
- Device Set: Ultrasonix L14-5 Ascenson
- Check if EM Sensors are in order: Transducer 1.: Probe (never unplug this), Transducer 2.: Reference, Sensor 1: Needle.
- Set parameters on US machine: Reject: 10, Gain: 75-90%, Frequency: 6.6M
- Open Plus Server Launcher on the computer
- C:/Users/Perklab/Documents/LumbarTutor/Config
- Device Set: Webcam Broadcasting
- Open Slicer 4.7.0 – 2017-05-24 on the computer
- Training Module – US Needle Tutor should start automatically
- Select Configuration: “Default” for the PT group, “No 3D guidance” for the Control group
- Change IP address to: 192.168.0.101
- Start Needle Tutor
-
Select Spine Model:
- Spine Model H02 for Practice 1
- Spine Model D01 for Practice 2
- Spine Model D03 for Final test
- Needle Calibration: complete pivot and spin calibration. The error for pivot calibration should be less than 0.5 mm.
When the participant arrives:
- Ask them if they watched the introductory presentation.
- Ask them if they have any question about the procedure or the study.
- Ask them to fill out the first part of the survey.
- Let them know which group they are (PT or Control) Remind the participants that no matter in which group she was selected, she will have access only for the U/S guidance during testing.
- Practice 1. Start the training on the first model. You might need to do hold the phantom in a stable position. The participants have 10 minutes for the first training. It can be finished earlier if the CSF flow occurs and the participant agrees.
- Practice 2. Change the spine model in the phantom. The second model is: PD01M01. Every time you change the model, you need to select the correct spine model in the LP Tutor!
- Don’t forget to fill the phantom with water and leave the valve open.
- Remind the participant that this is a very hard case, no need to be stressed and if the needle hit a bony structure pull the needle back, and try in a different direction. Do not force the needle. Again the outcome is getting the CSF Flow.
- Change the spine model in the phantom for the final test the model: PD02M01.
- Final test with using only US – no 3D view visible! Please do not answer questions during test.
- Click Start recording when participant start ultrasounding.
- Click Stop recording. End of procedure is when CSF flows back.
- Rename recordings: SubjectNumber_SpineModelNumber. You should have 3 recordings per person.
- Click Save all recordings.
- Delete all recordings
- Ask the participant to fill out the rest of the survey. Thank them for participation!
- and you are ready for the next participant.