Installation of firmware development environment is in another readme. This details the function and operation of the firmware. Wondering how the impedance measures are made? Want to get phase and magnitude out and play with that? Want to try different electrode stimulation patterns?
This firmware is meant for the Spectra: 32 electrode EIT device, which is also capable of bio-impedance spectroscopy as time series impedance measurements.
The device comes pre-programmed with this firmware. Each time you reset the device a set of menu options is communicated via UART. Once you select an option, press the return key and that option will start executing until you power cycle once more.
Options Outlines: A) Time series - This outputs the impedance magnitude at 40 frames per second at 50kHz(or whatever frequency is set in firmware).
How to take a measurement with bioimpedance spectroscopy: A+, V+, V- ,A-
B) Bioimpedance Spectroscopy - This outputs impedance magnitudes at the following frequencies, and then repeats the cycle. frequencies = {200,500,800,1000,2000,5000,8000,10000,15000,20000,30000,40000,50000,60000,70000};
Example of use:
C) Electrical Impedance Tomography - This is configurable in firmware to 8, 16, 32 electrodes dependent on desired timing and resolution trade offs etc.
The ordering of electrodes is created using pyEIT as follows: ex_mat = eit_scan_lines(ne=8, dist=3) step = 3 ex_mat = eit_scan_lines(ne=16, dist=8) step = 8 ex_mat = eit_scan_lines(ne=32, dist=16) step = 16 In the file the ordering is: A+, A-, V+, V- A, B, M, N where A : current driving electrode B : current sink M, N : boundary electrodes, where v_diff = v_n - v_m
We have chosen an opposition based electrode patten instead of the more commonly used adjacent scheme, as we came across these papers supporting improved contrast if opposition schemes were used: [1] A. Adler, P. O. Gaggero, and Y. Maimaitijiang, “Adjacent stimulation and measurement patterns considered harmful,” Physiol. Meas., vol. 32, no. 7, pp. 731–744, 2011. [2] F. M. Yu, C. N. Huang, F. M. Hsu, and H. Y. Chung, “Pseudo electrodes driven patterns for Electrical Impedance Tomography,” Proc. SICE Annu. Conf., pp. 2890–2894, 2007.
It's possible other electrode arrangement schemes are better yet.
Example of use:
A note on which debugger to use:
- I recommend using a Segger J-link debugger with SWD programming cable. If you are not using it for commercial use you can get the EDU version at a reasonable price(you don't even have to be part of a school). The configuration on the PCB also allows for the VCOM port to work which means you can also read serial through the same connection. The specific configuration can be seen in the .bat file contained in the EXE directory.