Expand Up @@ -71,15 +71,16 @@ If you enter carbs and select `Save without bolusing`, you have COB without asso

### When does Automatic Dosing Happen

Automatic dosing only happens when closed-loop is enabled in the settings screen.
Automatic dosing only happens when <code>Closed Loop</code> is enabled in the settings screen.

Loop generates a glucose prediction over the next 6 hours (the duration of insulin action), which is why the predicted glucose plot is included on the bolus screen. Loop considers glucose prediction with respect to your scheduled <code>Correction Range</code> over the full DIA, weighting closer predictions more than later predictions, when calculating [Recommended Bolus](../algorithm/bolus.md).

It is actually easier to answer when Loop will **not** automatically increase insulin delivery. In these cases, the prediction at the end of the DIA can be significantly higher than your <code>Correction Range</code> but no automatic dosing will happen:
It is actually easier to answer when Loop will **not** automatically increase insulin delivery.

* If at any time in the next 3 hours, Loop predicts glucose below <code>Glucose Safety Limit</code>, <code>Temp Basal</code> is immediately set to 0.0 U/hr and recommended bolus is set 0 U
* If the prediction dips below the low-end of your <code>Correction Range</code>, there is no **automatic** delivery
* If the current IOB is two times the <code>Maximum Bolus</code> setting, there is no **automatic** delivery
In the situations listed below, the prediction at the end of the DIA can be significantly higher than your <code>Correction Range</code> but no automatic increase in insulin delivery will occur:

* If at any time in the next 3 hours, Loop predicts glucose below <code>Glucose Safety Limit</code>, <code>Temp Basal</code> is immediately set to 0.0 U/hr and recommended bolus is set 0 U
* If the prediction dips below the low-end of your <code>Correction Range</code>, there is no **automatic** increase over scheduled basal
* If the current IOB is two times the <code>Maximum Bolus</code> setting, there is no **automatic** increase over scheduled basal

Even in cases where Loop does not automatically dose, the recommended bolus might be positive, which you see if you tap on bolus icon manually.
Even in cases where Loop does not automatically increase insulin delivery, the recommended bolus might be positive, which you see if you tap on the bolus icon manually.