thermohygrometer-calibration

Calibrate temperature and humidity sensors for ESPHome.

Using reference sensors alongside the sensors you want to calibrate, the tool will print out ESPHome-compatible YAML lines to calibrate the uncalibrated sensors.

This is assuming you already have reference sensors and that they have been placed next to the uncalibrated sensors. Wait a few days to gather enough data/variance for the tool to provide good calibration output.

You will not recognize the humidity calibration output because it is custom. I write about it, and a whole lot more about my adventure, here.

There are two entry points into this tool:

From InfluxDB

When provided necessary authentication details in an influx_config.py file you create, the tool is able to query the database directly to get the necessary values.

Example command:

python from_influx.py
    --start_time "2023-08-31 07:00:00" --end_time "2023-09-03 07:00:00"
    --reference_temperature_sensors sensor_a_temp,sensor_b_temp
    --reference_humidity_sensors sensor_a_hum,sensor_b_hum
    --uncalibrated_temperature_sensors sensor_c_temp,sensor_d_temp
    --uncalibrated_humidity_sensors sensor_c_hum,sensor_d_hum

This command treats "a" and "b" as reference sensors, ie. the ones we believe are reporting the correct values. No two sensors are ever identical, so the sensor values are averaged into a single "truth".

Sensors "c" and "d" are then compared against the reference "truth", and calibration YAML is outputted for each of these sensors.

The influx_config.py should contain this:

INFLUX_ORG='your_org'
INFLUX_BUCKET='your_bucket'  # probably "homeassistant"
INFLUX_TOKEN='your_token'
INFLUX_URL='http://your_db_ip:8086'

From CSV

If you already have the values on hand, you can pass them in using CSV files in the format:

SENSOR_ID,TIMESTAMP,VALUE

Which looks like this:

sensor_c_temp,2023-08-05T09:02:03Z,24.15

Example command:

python from_csv.py 
    --reference_temperature_csv ref_temp.csv
    --reference_humidity_csv ref_hum.csv
    --uncalibrated_temperature_csv uncal_temp.csv
    --uncalibrated_humidity_csv uncal_hum.csv

This route will behave the same as the Influx route. In fact, you can add --output_to_csv to the from_influx.py command and it will create the four CSVs that you can then pass into from_csv.py to output identical calibration configurations.

Example output

Sensor: sensor_c_temp

========== Temperature Calibration ==========
calibrate_linear:
  method: exact
  datapoints:
    - 21.985 -> 20.100
    - 23.958 -> 22.200
    - 25.942 -> 24.300
    - 27.808 -> 26.400
    - 29.633 -> 28.500
    - 31.490 -> 30.600
    - 33.600 -> 32.700
    - 35.722 -> 34.800

=========== Humidity Calibration ============
lambda: |-
  static auto expected1 = [](float x) -> float {
    return 38.065;
  };
  static auto expected2 = [](float x) -> float {
    return 48.819;
  };
  static auto measured1 = [](float x) -> float {
    static std::vector<std::vector<float>> mapping = {
      {23.116, 36.839}, {31.777, 38.114}
    };
    return segmented_linear(mapping, x);
  };
  static auto measured2 = [](float x) -> float {
    static std::vector<std::vector<float>> mapping = {
      {20.678, 44.844}, {28.367, 46.836}
    };
    return segmented_linear(mapping, x);
  };
  return calibrated_humidity(
    id(temperature).state,
    x, expected1, expected2, measured1, measured2
  );

The calibration.h file

You'll notice that the Humidity Calibration section calls segmented_linear() and calibrated_humidity(). These are not provided by ESPHome. I've provided these functions in to_esphome/calibration.h. Copy that file to your ESPHome instance and import it in your YAML like so:

esphome:
  name: my_sensor
  includes:
    - calibration.h

How to Install

Every platform is different so these instructions are only in general terms.

  1. Or install git and git clone this repository.
  2. Use a Python virtual environment with pip to set up the dependency for this tool.
    • Dependency installation command is typically: python3 -m pip install -r requirements.txt

Assumptions made

To avoid complicating the logic, I've made some assumptions about what's being provided to the tool.

  1. The set of sensors given for uncalibrated_temperature_sensors and uncalibrated_humidity_sensors are identical. Same length and contains the same sensors.
    • This means that your sensors are expected to be temperature+humidity combo sensors. This is true for me so I did not dive in further. If you are trying to calibrate sensors that are only one of these types... try to fake some dummy data. I would duplicate the data from another sensor and just change the name in the CSV, then use the from_csv.py route.
      • Make sure the dummy name matches the format of the sensor you want to calibrate, see the next assumption below.
    • Note that this restriction does not apply to the reference sensors. The tool will just average them together before using the data.
  2. Your sensor names are consistent between temperature and humidity. To match up the temperature and humidity entities for each sensor, I am assuming that the list of temperature entities and list of humidity entities would alpha-sort into the same order.
    • The respective sensors at each index of both lists should end up referring to the same actual sensor.