better specify LoC storing conditions #122

Open
soerenthomsen opened this issue Mar 3, 2022 · 1 comment
Open

better specify LoC storing conditions #122

soerenthomsen opened this issue Mar 3, 2022 · 1 comment

Comments

@soerenthomsen
Copy link
Member

soerenthomsen commented Mar 3, 2022

Anne comment:

"Which storage conditions? The NED reagent has to be stored at 4°C in dark and have a deadline of 1 month.
The reagents and working standards should be removed after each deployment to perform stability control measurements. The instrument itself should be stored without chemical (reagent, standards, Cd column). The luer lock syringe filter should also be removed and changed."

This relates to post recovery operations / storage.

@soerenthomsen
Copy link
Member Author

Following up to this

purchase vs. prepare reagents

"Or rather prepared… It is recommended that each user conduct its own studies to determine the stability of the reagents and working standards. Need a procedure or already provided by Clearwater?"